|
Emergency X-ray Generator
Shutdown |
- Push the red "Power
Off" button on the instrument cabinet. On the instruments with
four-circle goniometers, the needles on the kVolt and mAmpere meters
should begin moving to zero.
- Turn off electrical
power to the beige water chiller in the back corner of the lab.
- Close the yellow-handled
water valve above the water chiller.
back to top |
| Normal
X-ray Generator Shutdown |
- Decrease the current
and voltage until the minimum settings are reached. Alternate the current
and voltage adjustments. Allow at least 1 minute between adjustments.
- Older-style
generator (P31, P32)
- Simply turn
the knobs below the kV and mA meters counter clockwise.
- Newer-style
generator (P51)
- This is done
by holding Mode button down and hitting the down (or up) arrow until
either the "kV =" or "mA =" display appears;
then release the Mode button and hold the Edit button while hitting
the down arrow until the desired value is entered.
- Depress the X-ray
Off button.
- After 5 minutes
turn off the electrical power to the water chiller and close the water
valve to the city water supply
back to top |
| Starting
the X-ray Generator |
- Turn the valve
to the city water supply to a 45 degree angle to the water pipe.
- Turn on the electrical
power to the water chiller. The pump motor sounds should be audible.
Each instrument requires a flow rate of at least 1 gal / minute.
- Older-style
generator (P32, P31)
- Check
that the voltage and current adjustments are at their lowest
settings. (20 kV, 5 mA)
- Turn on
the power to the X-ray generator.
- Slowly
increase the voltage and current settings until the appropriate
values are reached. Alternate increases between voltage and
current. If the X-ray tube has not been used recently allow
at least 15 minutes between each voltage/current change. If
the tube has been used recently, 2 minutes should be sufficient
each between voltage/current change. If the needle on the current
meter fluctuates between adjustments, then the X-ray tube is
gassy, and the voltage/current changes should be made
with at least 10 minute intervals between adjustments. The product
of the final voltage and current should not exceed the maximum
power load marked on the top of the tube. The recommended voltage
and current values are shown on each X-ray generator.
- If
the X-ray generator will not turn on check the following:
- Hold
the X-ray On button in until the voltage needle increases
to 20 kV.
- If
the voltage meter does not change after 1 minute, there
is a serious problem with the generator--get help.
- If
the current meter reads 0 mA there is a serious problem
with the system--get help.
- If
the voltage meter returns to 0 kV when the X-ray On button
is released, there is a problem with the safety circuit.
Check that the X-ray On lamps on top of the tube and in
the cabinet door interlock are working. Check that the lamps
in the voltage and current meters are working. Check that
the spinner in the water flow meter next to the instrument
is spinning.
- If
the tube power, as shown on the meters, does not increase
to the values shown on the adjustment dials, there is probably
a problem with the voltage regulator--get help.
- Newer-style
generator (P51)
- Turn the
key on the generator control panel to the "I" position.
- On the
generator control panel, depress and hold the heater button
for at least 5 seconds or until the green ready lamp is lit.
The heater button is the black button between the Off and On
buttons.
- Push the
On button on the generator control panel. Note that the LCD
display should now read "kV = 20 ; mA = 5".
- Slowly
increase the voltage and current settings until the appropriate
values are reached. This is done by holding Mode button down
and hitting the down (or up) arrow until either the "kV
=" or "mA =" display appears; then release the
Mode button and hold the Edit button while hitting the up arrow
until the desired value is entered. Alternate increases between
voltage and current. If the X-ray tube has not been used recently
allow at least 15 minutes between each voltage or current change.
If the tube has been used recently, 2 minutes should be sufficient
between each voltage or current change. The product of the final
voltage and current should not exceed the maximum power load
marked on the top of the tube. The recommended voltage and current
values are shown on each X-ray generator.
back to top
|
| Starting
the Low-Temperature Device |
-
Turn on the gaseous
nitrogen flow to the low-temperature device. The nitrogen flow should
be set to about 12 SLPM for the four-circle goniometer instruments
and to about 20 SCFH for the platform goniometer instrument.
-
Fill the small
Dewar with liquid nitrogen. See specific instructions for filling
Dewars at http://cic.chem.wisc.edu/crystal/notes/LN2fill.html
-
Turn on power
to the low-temperature controller on top of the goniometer cabinet.
For the instruments with four-circle goniometers, adjust the nozzle
heater dials on the controller to the marked positions.
-
After about 30
minutes, the temperature should be stabilized, and the instrument
may be used. Be sure to follow the liquid nitrogen filling instructions
above to maintain the temperature during the experiment.
back to top |
| Shutting
Down the Low-Temperature Device |
-
Turn off power
to the low-temperature device controller. On the instruments with
four-circle goniometers, turn the dial for the nozzle heater counter
clockwise to its lowest setting.
- After two days,
turn off the flow of gaseous nitrogen to the low-temperature device.
back to top |
| Starting
the Diffractometer |
-
Turn on the electrical
power to the General Goniometer Control System, GGCS. Power to the
GGCS should remain on at all times when possible.
-
Turn on the electrical
power to the PC computer monitor and the computer itself. Power to
the computer and monitor should remain on whenever possible.
-
For the instruments
with charge-coupled device, ccd, detectors, turn on power to the supplemental
cooler and the camera electronics unit, CEU. The FTS cooler on the
instrument P32 takes about 30 minutes to achieve its running temperatures.
The Neslab cooler on the instrument P51 takes about 10 minutes to
achieve its running temperatures. When at proper running conditions,
the CEU on P32 should show that the temperature is about -55 degrees,
and the CEU on P51 should show that the temperature is about -45 degrees.
back to top |