CCD-Platform
Instructions
(Yzerman)
|
These instructions pertain primarily to
using the SMART CCD area detector with the Bruker Platform diffractometer
that is available at the University of Wisconsin, Chemistry Department.
-
Log into the NT computer (P51) next
to the instrument using the xtal account on the chmxray domain.
-
Start the SMART program on the PC
(Start, Programs, Bruker-AXS Programs, SMART).
-
Set up the project information for
this sample in Crystal/New Project. Both the working and data directories
should point to the same directory and should be on the o: device.
Be sure to include the chemical formula, crystal description, and
data collection temperature.
-
Optically align the sample using the
Goniometer/Optical option. Start the VIDEO program (Start, Programs,
Bruker-AXS Programs, Video) to view the crystal as it rotates. Be
sure that the crystal stays centered when rotated to all positions
using all four angle buttons of the manual control box.
-
Take a rotation frame image of the
crystal using the Acquire/Rotation option. If the rotation frame shows
no spots or if it shows powder diffraction rings, mount and align
another crystal.
-
Determine initial cell parameters
with the Crystal/Unit Cell command. Consider the following notes if
problems arise during the cell determination step.
-
If there are very few "usable" spots
in the cell parameter determination, when more were anticipated
from the rotation frame, then repeat the Crystal/Unit Cell step
with longer count times and a greater number of frames per run.
-
During the indexing step, you should
see one or more lines of "1"s in the output. If the program indexes
only with an implausibly large unit cell, try another crystal. If
the program has trouble indexing the spots or refining the cell
parameters and you feel this is otherwise a good crystal, collect
triclinic data and try the twinning programs.
-
Improve the cell parameters with the
following steps.
-
Select Crystal/Modify to use all
located spots.
-
Initially set the maximum deviation
RLV (Reciprocal Lattice Vector) to 0.05. In subsequent refinements,
use smaller values for the RLV deviations in order to remove any
large outliers. For the last refinement cycle, run Crystal/L S with
-1 for crystal system constraints.
-
Test the validity of the unit cell.
-
Check the fit of the spots to the
unit cell with Analyze/Display. Be sure to turn on the calculation
of hkl locations. Examine several frames to confirm that the cell
does properly fit the peak postions.
-
Estimate the volume of the molecule
(18 * # of nonhydrogen atoms). The estimated volume should be an
integer multiple of the cell volume.
-
Test to see if the structure has
been previously determined by checking the Crystal Data File Database.
This check is performed by the Crystal/CDF Search routine.
-
Optimize the scan runs using the ASTRO
program or use the following sets of runs. Note these are already
set up in the Acquire/EditHemi. Depending on the scattering ability
of your sample, you may want to change the count time.
# 2Theta Omega Phi Chi Axis Range # Time
1 -30 -30 0 54.9 2 -0.3 606 15
2 -30 -30 90 54.9 2 -0.3 435 15
3 -30 -30 180 54.9 2 -0.3 230 15
4 -30 -30 0 54.9 2 -0.3 50 15
-
Select the proper dark current correction
file for the desired scan time. The last three digits of the dark
current file names give the count times in seconds used to collect
the dark current data. If you can find no dark current file for your
count time or if you want to collect a newer dark current correction,
then go to the Detector/Dark Current menu and select the desired Seconds
per exposure and file name (E400L###._ls). Do not exceed 120 seconds
count time without discussing your data set with the lab director
-
Start data collection using either
the Acquire/Hemisphere or Acquire/MultiRun command. Be sure to set
the Job Name to the name being used for the project. Do not change
any other options except the Job Name or Title.
-
After data collection has completed,
start the data integration program SAINT Plus (Start, Programs, Bruker-AXS
Programs, SaintPLUS).
-
Set up project description by selecting
New under the Project menu. Enter the project name and indicate the
location of the data frames by selecting the .p4p file corresponding
to the first set of data runs.
-
Two additional steps are needed before
starting SAINT. In the SAINT menu, first select "Initialize". Then
under the SAINT menu select "Execute". In this display change the
"Crystal system constraint for integration(K1)" to "Triclinic". Check
the Bravais lattice type listed in the lower right corner of the screen.
Begin SAINT by selecting "Integrate, Sort, Global" button from the
SAINT menu.
-
When the integration is complete,
clean up the files with the following Command Prompt command: clean
newdir
-
Log out of the xtal account on P51.
Log onto one of the other NT computers in your own account and copy
the o:/newdir/newdir.* files to your directory. If the O: device is
not defined in your account, open Explorer (Start, Programs, Windows
NT Explorer), open the Map Network Drive icon, Select the drive letter
O: and map O: to the frames0 share on the P51 computer in the CHEMXRAY
domain.
back to top |
CCD-P4 Instructions
(Britney)
|
These
instructions pertain primarily to using a SMART CCD area detector with
the Bruker P4 diffractometer that is available at the University of Wisconsin,
Chemistry Department.
- Log into the NT
computer (P32) next to the instrument using the xtal account on the
chmxray domain.
- Start the SMART
program on the PC (Start, Programs, Bruker-AXS Programs, SMART). When
the program asks if the angles are at 0 degrees always say yes,
even if the angles are not at 0.
- Set up the project
information for this sample in Crystal/New Project. Both the working
and data directories should point to the same directory and should be
on the s:\ device. Be sure to include the chemical formula, crystal
description, and data collection temperature.
- Optically align
the sample using the Goniometer/Optical option. Drive all axes to zero
when done using Goniometer/Zero.
- Take a rotation
frame image of the crystal using the Acquire/Rotation option. If the
rotation frame shows no spots or if it shows powder diffraction rings,
mount and align another crystal.
- Determine initial
cell parameters with the following series commands.
- Test the validity
of the unit cell.
- Check the fit
of the spots to the unit cell with Analyze/Display. Be sure to turn
on the calculation of hkl locations. Examine several frames to confirm
that the cell does properly fit the peak postions.
- Estimate the
volume of the molecule (18 * # of nonhydrogen atoms). The estimated
volume should be an integer multiple of the cell volume.
- Test to see
if the structure has been previously determined by checking the Crystal
Data File Database. This check is performed by the Crystal/CDF Search
routine.
- Optimize the scan
runs using the ASTRO program or use one of the following sets of runs.
For triclinic
symmetry use the following runs. Note these are already set up in
the Acquire/EditHemi. Depending on the scattering ability of your
sample, you may want to change the count time.
# 2Theta Omega Phi Chi Axis Range # Time
1 30 30 0 40 3 0.4 910 20
2 30 30 0 15 3 0.4 910 20
3 30 -5 0 330 2 0.4 105 20
4 30 -5 90 330 2 0.4 105 20
5 30 -5 180 330 2 0.4 105 20
6 30 -5 270 330 2 0.4 105 20
7 30 30 0 40 3 0.4 50 20
For higher symmetry
samples the following series of runs usually proves adequate. These
runs are already entered in the Acquire/EditMulti set.
# 2Theta Omega Phi Chi Axis Range # Time
1 30 30 0 40 3 0.4 910 30
2 30 30 0 40 3 0.4 50 30
If greater redundancy
is needed, add more phi scans at different chi values(with our 30 deg
offset low temperature nozzle, chi must stay in the range -35 to +60
degrees). For weakly diffracting samples, increase the count times up
to a maximum of 120 sec/frame.
- Select the proper
dark current for the desired scan time in the Detector/Dark Current/Load
Dark menu. The last three digits of the dark current file names give
the count times in seconds used to collect the dark current data. If
you can find no dark current file for your count time or if you want
to collect a newer dark current correction, then go to the Detector/Dark
Current menu and select the desired Seconds per exposure and file name
(F404L###._dk).
- Start data collection
using either the Acquire/Hemisphere or Acquire/MultiRun command. Be
sure to set the Job Name to the name being used for the project. Do
not change any other options except the Job Name or the Title.
- After data collection
has completed, start the data integration program SAINT Plus (Start,
Programs, Bruker-AXS Programs, SaintPLUS).
- Set up project
description by selecting New under the Project menu. Enter the project
name and indicate the location of the data frames by selecting the .p4p
file corresponding to the first set of data runs.
- Two additional
steps are needed before starting SAINT. In the SAINT menu, first select
"Initialize". Then under the SAINT menu select "Execute". In this display
change the "Crystal system constraint for integration(K1)" to "Triclinic".
Check the Bravais lattice type listed in the lower right corner of the
screen. Begin SAINT by selecting "Integrate, Sort, Global" button from
the SAINT menu.
- When the integration
is complete, clean up the files with the following Command Prompt command: clean
newdir
- Log out of the
xtal account on P32. Log onto one of the other NT computers in your
own account and copy the s:/newdir/newdir.* files to your directory.
If the S: device is not defined in your account, open Explorer (Start,
Programs, Windows NT Explorer), open the Map Network Drive icon, Select
the drive letter S: and map S: to the frames4 share on the P32 computer
in the CHEMXRAY domain.
|
Filling Liquid Nitrogen
Dewars
|
-
Filling the
Small Dewar From the Large Dewar.
The
small Dewar on P32 needs to be filled at least once per day. The Dewar
connected to P51 needs to be filled at least twice per day (suggest
morning and evening).
Each
large Dewar should have an adjustable wrench and a canvas glove. The
glove should be used as a "hot pad" to insulate your hand from the very
cold metal valves and pipes.
-
Connecting
Dewars. Thread the brass fitting on the flexible metal hose
from the small Dewar to the "liquid" port of the large Dewar by
hand. Finish tightening the brass fitting with the adjustable wrench.
-
Filling
the small Dewar. Open the valve on the "liquid" port of the
large Dewar until a pulsing sound is heard. After about 1-2 minutes
the pulsing sound should stop. Now open the valve on the "liquid"
port all the way open using the cloth glove. Once the valve is opened,
place the glove on top of the blue level indicator on the large
Dewar.
Note
that the small Dewars typically take about 10-20 minutes to fill.
Do not leave the room when filling the small Dewars, or it is likely
that the small Dewar will overflow and the entire contents of the
large Dewar will empty onto the floor.
-
Slow
filling. If nitrogen is flowing slowly from the large Dewar,
as often happens after the large Dewar has been filled, then open
the "Pressure Building" valve on top of the large Dewar two full
turns.
-
Whistling
sound. If you hear a whistling sound coming from the large Dewar,
then the large Dewar is empty. Close all valves on top of the large
Dewar using the cloth glove, and refill according to the instructions
below.
-
Completing
the fill. When the small Dewar is full, close the valve on the
"liquid" port of the large Dewar using the cloth glove. If the "Pressure
Building" valve was opened during the fill, be sure to close this
valve also. Each day record the group or groups to be charged for
the nitrogen and the date on the log sheet attached to the Dewar.
-
Filling the
Large Dewar.
The
large Dewars have a float-type indicator, which is covered in blue plastic,
that shows the approximate level of liquid nitrogen in the tank. When
this marker is at the bottom, or when a whistling sound is heard during
a fill of the small Dewar, then the large Dewar needs to be refilled.
-
Disconnect
from the small Dewar. Close all valves on top of the large Dewar
using the cloth glove. Use the wrench to loosen and disconnect the
flexible hose at the "liquid" port of the large Dewar. On the P51
instrument place the end of the hose on the support attached to
the instrument shell.
-
Connect
the large Dewar to the filling station.
- Pull the
large Dewar to the filling station located in room B371B.
- Connect
the "vent" port to the flexible hose on the right side of the
duct work by hand.
- Connect
the "liquid" port to the flexible hose on the left side of the
duct work by hand.
- Using the
wrench, tighten only the tubing connected to the "liquid" port.
- Open the
"liquid" valve and the manual valve on the manifold next to the
"liquid" port connection.
- Set the
timer on the wall for 25 minutes and push the red button.
- Open the
"vent" valve of the large Dewar.
-
Record
the nitrogen used. Divide the number of days that the large
Dewar was used into 160 liters. Record the group name and proportionate
amounts of liquid nitrogen used on the log sheet located to the
left of the timer. Record the "student" simply as "x-ray". Place
a tick mark on the log sheet attached to the Dewar to indicate when
the Dewar was filled.
-
Completing
the fill.
- When the
large Dewar is full, if the timer arrow is not reset to the 25
minutes, turn the arrow counter-clock wise until the timer stops
and resets itself to 25 minutes.
- Close all
manual valves on the Dewar using the cloth glove.
- Close the
manual valve on the manifold that was attached to the "liquid"
line.
- Using the
wrench disconnect the two flexible hoses from the large Dewar.
Use the cloth glove to support the flexible hoses while disconnecting.
Note, if the fittings are frozen to the short pipes, then use
the heat gun located to the right of the flexible hoses to thaw
out the frozen fittings.
- Place the
glove on top of the blue plastic indicator, and place the wrench
on top of the large Dewar.
- Return the
large Dewar to the instrument room. Be sure to relock any doors
that were unlocked when taking the Dewar to be filled.
back
to top |
Retrieving Data
Files
|
Data files
for completed crystal structures are stored on the web server cic.chem.wisc.edu.
Files may be retrieved from the server by using web browser. Simply point
to the url http://cic.chem.wisc.edu/crystal/result.html
and then select the proper research group and project. At this point you
will be asked for a username and password to access the data files.back
to top |
| |
|
| |
|
| |
|
| |
|
| |
|
| |
|